Contrôle Analytique's valve port system and purge/sealing groove system is so flexible that it can be used in many configurations.
All these configurations could be actuated by an independently actuated plunger on each port.
-
KDV-6
6 Port Selection Valve
This configuration could be used for direct replacement of
existing old type diaphragm valves. Typical sample injection
like shown in fig. 5 and fig. 6 could be realized.
-
KDV-4
4 Port Selection Valve
This 4 port configuration allows many switching flow paths, and this with no dead volume effect.
-
KDV-S
Sample Stream Selection Valve
Each port being independently actuated, sample stream flow
path scheme or multiple column switching could be easily
done. The sealing/purging groove prevents any inboard/outboard
contamination and allows the possibility of issuing a
proper alarm in case of leak.
|
New Chromatographic Method
This unique design allow for 2 new gas chromatographic method, like:
- Sample loop pressurization before injection (3-step sample injection) eliminating flow and pressure surge upon injection.
- Peak slicing, allow easy heart cutting and individual peak concentration, increasing at least by two the system sensitivity.
For example, application like the measurement of O2 and H2 become very easy. Ask for application bulletin, for KDV valve series application.
The peak slicing method is also covered by an international (PCT) patent pending.
See also product brochure and application bulletin for our new rotary g.c. valve design.