 Nitrogen in Argon or Helium  on line Gas Analyzer



Enlarge Product Image
Technical Data
Sheet (556 Kb)
Features
  • Display resolution to 10 ppt
  • Plasma emission system
  • No sensor or detector to replace,maintenance free
  • Microprocessor controlled
  • Low sample flow
  • Fast response
Applications
  • Electronic wafer fabs
  • Gas management system
  • Gas purifier efficiency monitoring
  • Specialty gas laboratories
  • High purity Argon plant
Specifications
Detector type Single beam. Electromagnetically induced plasma cell. Non ratiometric measurement, Material : Quartz, single element, vacuum tight to 10 PSIG (69 KPAG)
Range
  • X1 : 0 - 100 ppb display resolution to 10 ppt
  • X10 : 0 - 500 ppb display resolution to .1 ppb
  • X100 : 0 -5 ppm display resolution to 0.01 ppm
Accuracy
  • ± 1.5% (or better) of reading on 0 -100 / 0 - 500 ppb
  • ± 1% on 5 ppm range
Sensitivity Guarantee 1 ppb
Sample flow requirements 25 sccm
Recommended calibration gas
  • Zero : from purifier
  • Span : 1 to 3 ppm or dilution system (orifice based)
Flow accuracy 0 to 200 cc ± 1 % full scale
Minimum input pressure 8 PSIG, for 25 cc/min (55 KPAG)
Recommended operating pressure 10 PSIG to 20 PSIG for 25 cc/min (70 KPAG to 140 KPAG)
Operating temperature 10° C to 35° C (software temperature compensation). Detector system is temperature regulated in an insulated compartment (thermo-cool system)
Sample vent and by-pass vent 1/8" Swagelok
Simple inlet 1/8" VCR
Supply 115 VAC, 50 - 60 Hz or 220 VAC, 50 - 60 Hz (on order)
Power consumption Maximum 35 watts
4 - 20 mA
  • 4 - 20 mA isolated output
  • Maximum load impedance : 600 ohm at 20 mA DC
  • Insulation : 750 VDC or AC peak maximum
Response time Less than 90 seconds for 63% step change at 25 cc (tipically 60 seconds)
Weight 15 kg (33 lbs)
Standard features
  • Fully microprocessor-controlled, with 4 X 40 LCD display
  • Self-diagnostic system software
  • Digital output : 4 digitals outputs. 3 are for range indication and 1 for system status monitoring
  • Two alarms contact output
  • Isolated 4 - 20 mA output
Options
  • Serial port : RS-232
  • Custom options : contact factory
Dimensions
  High (mm) Deep (mm) Wide (mm)
Standard rack mount 3U 5.25" (133.3) 18" (457.0) 19" (482.6)