| Detector Type |
Electromagnetically induced plasma cell (dual beam) |
| Detector Material |
Quartz, single element, vacuum tight to 10 PSI (69kPA)/New Patent Pending Detector |
| Ranges |
- 0-1 ppm, resolution of 10 ppb, user configurable up to 10000 ppm
- 0-10 ppm, resolution of 0.1 ppm, user configurable up to 10000 ppm
- 0-100 ppm, resolution of 1 ppm, user configurable up to 10000 ppm
|
| Accuracy |
Better than ± 1% full scale |
| Sample Flow Requirements |
20 to 150 sccm |
| Gas Connections |
Sample: 1/8" Swagelok or 1/8" VCR tube fitting |
| Vent |
1/8" swagelock tube fitting |
| Recommended Calibration Gas |
Zero: GP-200 purified gas (Getter) |
| Flow Accuracy |
0 to 200 sccm ± 1% full scale |
Recommended Maximum Operating Pressure |
20 PSIG (138 kPAG) |
Recommended Minimum Operating Pressure |
4 PSIG (28 kPAG) |
| Operating Temperature |
10 °C to 35 °C |
| Supply |
115 VAC, 50 – 60 Hz or 220 VAC, 50 – 60 Hz |
| Power Consumption |
35 watts (maximum) |
| Response Time |
20 seconds for 90% of a step change at 75 sccm (typically 10 seconds) |
| Drift |
± 1% over 24 hours (typically ± .5% over long term) |
| Weight |
15 kg (33lbs) |
| Connectors |
- RJ-45 for Ethernet 10Base-T
- DB-9 for RS-232 communication
- Curtis GTFX-10 screw terminal block for I/Os
- Corcom 5EFM4S power entry module
|
| Standard Features |
- Microprocessor controlled with 4x40 LCD display and a 4x4 keypad
- Self diagnostic system software
- Manual or auto-ranging (user selectable)
- Internet Web Enable
- Real Time Clock
- Alarm Historic
- CE conform and CSA on demand
- Digital ouputs for remote monitoring:
(all dry relay contacts, SPST 1 amp 250 Vac resistive load)
- System status (1 output)
- Range in use (3 output)
- 2 alarm outputs (user programmable set point)
|
| Options |
- Auto-zero and auto-span calibration at user defined intervals
- 4-20 mA isolated output:
- 600 ohms at 20 mA dc
- Fuse protection
- Transient protected
- Isolated from power ground and analyzer's electrical circuit
- Programmable for over- and under-drive in case of a system alarm
- Manual software control for calibration, monitoring and troubleshooting
- Isolation: 750 VDC or AC peak maximum
- Serial port: RS-232/422/485/modem
- Custom option: contact factory
|